Remote Manual SIGLENT
SDM3055/SDM3055A 58
2.9.4 [SENSe:]CURRent:{AC|DC}:RANGe {<range>|MIN|MAX|DEF}
[SENSe:]CURRent:{AC|DC}:RANGe? [{MIN|MAX|DEF}]
Manually choose measurement range.
Note:This parameter is not shared between AC and DC measurements. The parameters
are independent for AC and DC measurements.
Parameter Typical Return
<range>{200uA|2mA|20mA|200mA|2A|10A|AUTO}
default is AUTO(autorange).
+2.00000000E-01
Configure AC current measurements using the 2 A range. Make and read two
measurements:
CONF:CURR:AC
CURR:AC:RANG 2
SAMP:COUN 2
READ?
Typical Response +3.53049833E-04,+3.54828343E-04
Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging.
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns "9.9E37" from the
remote interface.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument
Preset (SYSTem:PRESet).