SIGLENT Remote Manual
37 SDM3055/SDM3055A
2.6.1 CONFigure?
Returns a quoted string indicating the present function, range, and resolution. The short
form of the function name (CURR:AC, FREQ) is always returned.
Parameter Typical Return
(none) “VOLT +2.00000000E-01
Return the present function, range, and resolution.
CONF?
2.6.2 CONFigure:CONTinuity
Sets all measurement parameters and trigger parameters to their default values for
continuity measurements.
Parameter Typical Return
(none) (none)
Configure the instrument for continuity measurements.Then make a measurement using an
external trigger with positive slope (rising edge) and read the measurement:
CONF:CONT
TRIG:SOUR EXT;SLOP POS
READ?
The range and resolution are fixed at 1 for continuity tests (a 2-wire resistance
measurement).
The instrument beeps (if the beeper is enabled) for each measurement less than or
equal to the continuity threshold (The threshold can be passed by [: SENSe]:
CONTinuity:THReshold:VALue command ), and the actual resistance measurement
appears on the display.
From 10 Ω to 1.2 kΩ, the instrument displays the actual resistance measurement
with no beep. Above 1.2 kΩ,the instrument displays "OPEN" with no beep.
The FETCh?, READ?, and MEASure:CONTinuity? queries return the measured
resistance, regardless of its value.
Use READ? or INITiate to start the measurement.