SIGLENT Remote Manual
11 SDM3055/SDM3055A
1.6. READ?
Read and erase measurement results from reading storage ,up to the specified <
max_readings >.This command reads the data will begin with the first measurement
results.
Parameter Typical Return
(none)
-1.23006735E-03,-1.30991641E-03,-1.32756530E-03,
-1.32002814E-03
Transmission measurements from the reading memory:
TRIG:COUN 4
SAMP:COUN 1
READ?
The FETch? query does not erase measurements from the reading memory. You can
send the query multiple times to retrieve the same data.
You can store up to 1,000 measurements in the reading memory of the SDM3055x. If
reading memory overflows, new measurements will overwrite the oldest
measurements stored; the most recent measurements are always preserved. No
error is generated, but the Reading Mem Ovfl bit (bit 14) is set in the Questionable
Data Register's condition register .
It is important to note that the following command will measurements to empty,
leading to FETCh? The return value of the change:
INITiate
MEASure:<function>?
READ?
*RST
SYSTem:PRESet