Remote Manual SIGLENT
SDM3055/SDM3055A 50
2.8.4 MEASure:{FREQuency|PERiod}?
Sets all measurement parameters and trigger parameters to their default values for
frequency or period measurements and immediately triggers a measurement. The results
are sent directly to the instrument's output buffer.
Parameter Typical Return
(none) +7.19480528E+01
configure and read the default frequency measurements:
MEAS:FREQ?
If the input voltage is too large for the selected voltage range (manual ranging), the
instrument displays the word Overload on the front panel and returns "9.9E37" from
the remote interface. Autoranging can be enabled for the input voltage.
2.8.5 MEASure:{RESistance|FRESistance}?
[{<range>|AUTO|MIN|MAX|DEF}]
Sets all measurement and trigger parameters to their default values for 4-wire resistance
(FRESistance) or 2-wire (RESistance) measurements, and immediately triggers a
measurement. The results are sent directly to the instrument's output buffer.Also specifies
the stalls through the incoming parameters.
Parameter Typical Return
<range>:{200 Ω|2 kΩ|20 kΩ|200 kΩ|2 MΩ|10
MΩ|100 MΩ},AUTO
(default) or DEFault
(none)
Configure 4-wire resistance measurements using the 200 Ω range with
default resolution. Then make and read one measurement::
MEAS:FRES? 200
Typical Response: +6.71881065E+01