47 SDM3055/SDM3055A
2.8.1 MEASure:CONTinuity?
Sets all measurement parameters and trigger parameters to their default values for
continuity test and immediately triggers a measurement. The results are sent directly to
the instrument's output buffer.
Parameter Typical Return
none +9.84739065E+02
Configure the instrument for continuity measurements. Then make and read one
measurement:
MEAS:CONT?
The range and resolution are fixed at 1 for continuity tests (a 2-wire resistance
measurement).
The instrument beeps (if the beeper is enabled) for each measurement less than or
equal to the continuity threshold (≤10 Ω), and the actual resistance measurement
appears on the display.
From 10 Ω to 1.2 kΩ, the instrument displays the actual resistance measurement with
no beep. Above 1.2 kΩ, the instrument displays "OPEN" with no beep.
The FETCh?, READ?, and MEASure:CONTinuity? queries return the measured
resistance, regardless of its value.