Remote Manual SIGLENT
SDM3055/SDM3055A 40
2.6.6 CONFigure:{RESistance|FRESistance}
[{<range>|AUTO|MIN|MAX|DEF}]
Sets all measurement parameters and trigger parameters to their default values for 4-wire
(FRESistance) or 2-wire (RESistance) resistance measurements.Also specifies the range
and resolution.
Parameter Typical Return
<range> {200 Ω|2 kΩ|20 kΩ|200 kΩ|2 MΩ|10
MΩ|100 MΩ}AUTO(default)
none
Configure 4-wire resistance measurements using the 200 Ω range with default
resolution.Make and read two measurementsCONF:FRES 200
SAMP:COUN 2
READ?
Typical Response +6.71881065E+01,+6.83543086E+01
You can let autoranging select the measurement range, or you can manually select a
fixed range. Autoranging conveniently selects the range for each measurement
based on the input signal. For fastest measurements, use manual ranging
(autoranging may require additional time for range selection).
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns "9.9E37" from the
remote interface.
Use READ? or INITiate to start the measurement.