2.6.4 CONFigure:DIODe
Sets all measurement parameters and trigger parameters to their default values for diode
tests.
Parameter Typical Return
none none
Configure, make, and read a default diode
measurement:
CONF:DIOD
READ?
Typical Response +1.32130000E-01
The range and resolution are fixed for diode tests: the range is 1 VDC.
The FETCh?, READ?, and MEASure:DIODe? queries return the measured voltage,
regardless of its value.
Use READ? or INITiate to start the measurement.
2.6.5 CONFigure:{FREQuency|PERiod}
Sets all measurement parameters and trigger parameters to their default values in the
frequency/cycle mode.
Parameter Typical Return
(none) none
Configure, make, and read a default
frequency measurement
CONF:FREQ
READ?
Typical Response +7.79645018E+01
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns
"+9.90000000E+37" from the remote interface.