SIGLENT Remote Manual
45 SDM3055/SDM3055A
2.7.2 DATA: POINts?
Returns the total number of measurements currently in reading memory. You can execute
this query at any time, even during a series of measurements.
Parameter Typical Return
none +20
Return the number of measurements in
reading memory:DATA:POIN?DATA:POIN?
You can store up to 1,000 measurements in the reading memory on the SDM3055x.
2.7.3 DATA:REMove?<num_readings> [,WAIT]
Reads and erases <num_readings> measurements from the reading memory. If fewer
than <num_readings> measurements are available, the query will return an error unless
the WAIT parameter is specified, in which case the query will wait until <num_readings
measurements are available.
Parameter Typical Return
1~10000
-4.55379486E-04,-4.55975533E-04,-4.56273556E-04,
-4.53591347E-04,-4.55379486E-04
Read and erase the five oldest readings from reading memory:
DATA:REMove? 5
The R? and DATA:REMove? queries allow you to periodically remove measurements
from the reading memory that would normally cause the reading memory to overflow.