SIGLENT SDM3055 Bedienungsanleitung Seite 28

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 97
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 27
SIGLENT Remote Manual
27 SDM3055/SDM3055A
2.4.2 CALCulate:SCALe:DBM:REFerence {<reference>|MIN|MAX|DEF}
CALCulate:SCALe:DBM:REFerence? [{MIN|MAX}]
Selects the reference resistance for converting voltage measurements to dBm. This
reference value affects the dBm and dB scaling functions.
Note:This command applies only to ACV and DCV measurement functions.
Parameter Typical Return
507593110124125135
150250300500600((default)
80090010001200 or 8000Ω
+6.00000000E+02
Enable dBm scaling with a reference resistance of 600 Ω:
CALC:SCAL:DBM:REF 600
CALC:SCAL:FUNC DBM
CALC:SCAL:STAT ON
The instrument sets the reference value to its default value after a Factory Reset
(*RST), an Instrument Preset (SYSTem:PRESet), a change in math function, or a
change in measurement function.
Seitenansicht 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 96 97

Kommentare zu diesen Handbüchern

Keine Kommentare