Remote Manual SIGLENT
SDM3055/SDM3055A 26
2.4 CALCulate:SCALe Subsystem
This subsystem scales ACV and DCV measurements.
Command Summary
CALCulate:SCALe:DB:REFerence
CALCulate:SCALe:DBM:REFerence
CALCulate:SCALe:FUNCtion
CALCulate:SCALe:REFerence:AUTO
CALCulate:SCALe[:STATe]
2.4.1 CALCulate:SCALe:DB:REFerence {<reference>|MIN|MAX|DEF}
CALCulate:SCALe:DB:REFerence? [{MIN|MAX}]
Stores a relative value in the multimeter's dB Relative Register, which is used for the dB
function in CALCulate:SCALe:FUNCtion. When the dB function is enabled, this value will
be subtracted from the each voltage measurement after the measurement is converted to
dBm.
Note: This command applies only to ACV and DCV measurement functions.
Parameter Typical Return
-200.0 dBm to +200.0 dBm, default 0
+5.00000000E+02
Enable dB scaling with a -10 dB reference reference resistance:
CALC:SCAL:DB:REF -10.0
CALC:SCAL:FUNC DB
CALC:SCAL:STAT ON
Specifying a reference value disables automatic reference selection
(CALCulate:SCALe:REFerence:AUTO OFF).
The dB relative value parameter is relative to the dBm reference set with
CALCulate:SCALe:DBM:REFerence.
The instrument sets the reference value to 0.0 with automatic reference selection
enabled after a Factory Reset (*RST), an Instrument Preset (SYSTem:PRESet), a
change in math function, or a change in measurement function.