Remote Manual SIGLENT
SDM3055/SDM3055A 84
[SENSe:]CAPacitance Subsystem
This subsystem is configured capacitance measurement.
.
Command Summary
[SENSe:]CAPacitance:NULL[:STATe]
[SENSe:]CAPacitance:NULL:VALue
[SENSe:]CAPacitance:NULL:VALue:AUTO
[SENSe:]CAPacitance:RANGe
[SENSe:]CAPacitance:RANGe:AUTO
2.9.33 [SENSe:]CAPacitance:NULL[:STATe] {ON|1|OFF|0}
[SENSe:]CAPacitance:NULL[:STATe]?
Enable or disable the zero function.
Parameter Typical Return
{ON|1|OFF|0}default ON 0(OFF)or1(ON)
Use zero functions from the measured value minus 100 nf to configure the
capacitance measurement.Make and read two measurements,sent the results
to the output buffer of the instrument.
CONF:CAP
CAP:NULL:STAT ON;VAL 100nF
SAMP:COUN 2
READ?
Typical Response: +4.79899595E-10,+4.79906446E-10
Enable the scaling function will also enable automatic zero value
choice.([SENSe:]Capacitance:{AC|DC}:NULL:VALue:AUTO ON).
To set a fixedr eturn null,use this
command:[SENSe:]Capacitance:{AC|DC}:NULL:VALue.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument
Preset (SYSTem:PRESet).