SIGLENT SDM3055 Bedienungsanleitung Seite 78

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 97
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 77
SIGLENT Remote Manual
77 SDM3055/SDM3055A
[SENSe:]VOLTage Subsystem
This subsystem configures AC voltage measurements and DC voltage and ratio
measurements.
Command Summary
[SENSe:]VOLTage:{AC|DC}:NULL[:STATe]
[SENSe:]VOLTage:{AC|DC}:NULL:VALue
[SENSe:]VOLTage:{AC|DC}:NULL:VALue:AUTO
[SENSe:]VOLTage:{AC|DC}:RANGe
[SENSe:]VOLTage:{AC|DC}:RANGe:AUTO
[SENSe:]VOLTage[:DC]:NPLC
[SENSe]:VOLTage[:DC]:FILTer[:STATe]
[SENSe:]VOLTage[:DC]:IMPedance
2.9.25 [SENSe:] VOLTage:{AC|DC}:NULL[:STATe] {ON|1|OFF|0}
[SENSe:] VOLTage:{AC|DC}:NULL[:STATe]?
Enables or disables the null function for AC or DC voltage measurements.
Note:This parameter is not shared between AC and DC measurements. The parameters
are independent for ACand DC measurements.
Parameter Typical Return
{ON|1|OFF|0}default OFF 0(OFF)or1(ON)
Configure AC voltage measurements, using the null function to subtract 100
mV from the measurements. Make and read two measurements:
CONF:VOLT:AC
VOLT:AC:NULL:STAT ON;VAL 100mV
SAMP:COUN 2
READ?
Typical Response +1.03625390E+00,+1.03641200E+00
Enabling the scaling function also enables automatic null value selection
([SENSe:]VOLTage: {AC|DC}:NULL:VALue:AUTO ON).
To set a fixed null value, use this command:
[SENSe:]VOLTage:{AC|DC}:NULL:VALue.
The instrument disables the null function after a Factory Reset (*RST), Instrument
Preset (SYSTem:PRESet), or CONFigure function.
Seitenansicht 77
1 2 ... 73 74 75 76 77 78 79 80 81 82 83 ... 96 97

Kommentare zu diesen Handbüchern

Keine Kommentare