SIGLENT Remote Manual
71 SDM3055/SDM3055A
2.9.18 [SENSe:]{RESistance|FRESistance}:RANGe:AUTO
{OFF|ON|ONCE}
[SENSe:]{RESistance|FRESistance}:RANGe:AUTO?
Disables or enables autoranging for all resistance measurements. Autoranging is
convenient because it automatically selects the range for each measurement based on
the input signal. Specifying ONCE performs an immediate autorange and then turns
autoranging off.
Parameter Typical Return
{ON|1|OFF|0}default ON 0(OFF) or 1(ON)
Configure 2-wire resistance measurements and perform an immediate
autorange. Make and read two measurements:
CONF:RES
RES:RANG:AUTO ONCE
SAMP:COUN 2
READ?
Autoranging goes down a range at less than 10% of range and up a range at greater
than 120% of range.
Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument
Preset (SYSTem:PRESet).