67 SDM3055/SDM3055A
2.9.14 [SENSe:]{RESistance|FRESistance}:NULL[:STATe] {ON|1|OFF|0}
[SENSe:]{RESistance|FRESistance}:NULL[:STATe]?
Enables or disables the null function for all resistance measurements.
Parameter Typical Return
{ON|1|OFF|0},default OFF 0 (OFF) or 1 (ON)
Configure 2-wire resistance measurements, using the null function to remove
100 mΩ of wiring resistance. Make and
read two measurements
CONF:RES
RES:NULL:STAT ON;VAL 0.1
SAMP:COUN 2
READ?
Enabling the scaling function also enables automatic null value selection
([SENSe:]{RESistance/FRESistance}:NULL:VALue:AUTO ON).
To set a fixed null value, use this command:
[SENSe:]{RESistance/FRESistance}:NULL:VALue.
The instrument disables the null function after a Factory Reset (*RST), Instrument
Preset (SYSTem:PRESet), or CONFigure function.