Remote Manual SIGLENT
SDM3055/SDM3055A 62
2.9.9 [SENSe:]{FREQuency|PERiod}:NULL:VALue
{<value>|MIN|MAX|DEF}
[SENSe:]{FREQuency|PERiod}:NULL:VALue? [{MIN|MAX|DEF}]
Set to zero
This parameter is shared between frequency and period measurements.
Parameter Typical Return
-1.2E6 to +1.2E6, default 0 +1.00000000E-02
Configure frequency measurements, using the null function to subtract 1kHz
from the measurements. Make and read two measurements:
FREQ:NULL:STAT ON;VAL 1 kHz
SAMP:COUN 2
READ?
Typical Response: +1.15953012E+02,+1.07954466E+02
Specifying a null value disables automatic null value selection
([SENSe:]{FREQuency|PERiod}:NULL:VALue:AUTO OFF).
To use the null value, the null state must be on
([SENSe:]{FREQuency|PERiod}:NULL:STATe ON).
This parameter is set to its default value after a Factory Reset (*RST), Instrument
Preset (SYSTem:PRESet), or CONFigure function.
Kommentare zu diesen Handbüchern