Remote Manual SIGLENT
SDM3055/SDM3055A 52
2.8.7 MEASure[:VOLTage]:{AC|DC}? [{<range>|AUTO|MIN|MAX|DEF}]
Sets all measurement parameters and trigger parameters to their default values and
immediately triggers a measurement. The results are sent directly to the instrument's
output buffer.Also specifies the stalls through the incoming parameters.
Parameter Typical Return
<range>:{200 mV|2 V|20 V|200 V|1000
V(DC)/750V(AC)},default AUTO (automatic adjustment
range)
(none)
Configure AC voltage measurements using the 200 V range. Then make and read
one measurement:
MEAS:VOLT:AC? 200
Typical Response: +2.43186951E-02,+2.56896019E-02
You can let autoranging select the measurement range, or you can manually select a
fixed range. Autoranging conveniently selects the range for each measurement
based on the input signal. For fastest measurements, use manual ranging
(autoranging may require additional time for range selection).
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns "9.9E37" from the
remote interface.