SIGLENT SDM3055 Bedienungsanleitung Seite 23

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 97
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 22
Remote Manual SIGLENT
SDM3055/SDM3055A 22
2.3.4 CALCulate:TRANsform:HISTogram:POINts{<value>|MIN|MAX|DEF}
CALCulate:TRANsform:HISTogram:POINts?[{MIN|MAX|DEF}]
Sets the number of bins between the lower and upper range values for the histogram. Two
additional bins always exist:one for measurements below the lower range and one for
measurements above the upper range.
Parameter
Typical
Return
{10|20|40|100|200|400|MIN|MAX|DEF},default
100 +100
See Example.
You can specify the lower and upper range values using
CALCulate:TRANsform:HISTogram:RANGe: {LOWer|UPPer}. Lower and upper
range values are computed automatically if
CALCulate:TRANsform:HISTogram:RANGe:AUTO is ON.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument
Preset (SYSTem:PRESet).