Remote Manual SIGLENT
SDM3055/SDM3055A 12
1.7. SAMPle:COUNt {<count>|MIN|MAX|DEF}
SAMPle:COUNt? [{MIN|MAX|DEF}]
Specifies the number of measurements (samples) the instrument will take per trigger.
Parameter Typical Return
1( default) 100,000 1
Set a single trigger sampling frequency for 10 times, the back
panel of the VM Comp BNC connector on the oscilloscope will only
observe a pulse:
SAMP:COUN 10 //Set the sampling frequency for 10
times
TRIG:COUN 1 //Set the trigger for 1 times
TRIG:SOUR EXT;SLOP NEG //Set the trigger source to the
external trigger and trigger signal
is set to the falling edge
OUTP:TRIG:SLOP POS //Set the trigger output signal to rise
READ? //Start the wheel measurement and
reading
You can use the specified sample count in conjunction with a trigger count
(TRIGger:COUNt), which sets the number of triggers to be accepted before returning
to the "idle" trigger state. The total number of measurements returned will be the
product of the sample count and trigger count.
You can store up to 1,000 measurements in the reading memory of the SDM3055x. If
reading memory overflows, new measurements will overwrite the oldest
measurements stored; the most recent measurements are always preserved. No
error is generated, but the Reading Mem Ovfl bit (bit 14) is set in the Questionable
Data Register's condition register .
This parameter is set to its default value after a Factory Reset (*RST) or Instrument
Preset (SYSTem:PRESet).