I User Manual SDS1000 Series Digital Storage Oscilloscope Version No.: V 1.3 Si
2 2.13.4 Self Calibration...102 2.13.5 Self Test...
92 2.13 Utility System “UTILITY”: The Utility System Button. Utility Setup button Picture 2.13-1 Table 2-48 Utility System f
93 Picture 2.13-2 Table
94 Table 2-50 Functional Menu of the Utility System 3: Option Setting Introduction Update firmware You can update the o
95
96 have a higher request to the cable , to electricity characteristic of the port. When Higher Speed connect fail, you can try to return to
97 2.13.1 System status Pre
98 Picture 2.13-7 2.13.3 Print The scope support PictBridge compatible printers. You can connect the back USB port of the oscilloscope to the US
99 Picture 2.13-8 Tabl
100 Picture2.13-9 To print screen picture, follow these steps: 1. Connect the oscilloscope to a PictBridge compatible printer. 1) I
101 settings of your printe
102 2.13.4 Self Calibration Self Calibration procedure can optimize the signal path at the most measurement precision..You can run this procedure
103 “Single” for test.
104 Picture 2.13-12 3. LED test ● Select “LED Test” to enter the lighten interface, the on-screen lathy rectan
105 2.13.6 Updating the Sys
106 9. Gate Way:10.11.0.1 10. Press‘Save/Recall’button and ‘single’button 11. name the correlated model upgrade program as ‘Upgrade.ads’,and add
107 Picture 2.13-15 Table
108 Table 2-58 Mask Setting function menu 1: Option Setting Instruction X Mask xdiv Turn the “Universal” knob to set the range of horizo
109 Picture 2.13-18
110 option button to set the output option. 13) Enter the first page of “Pass/Fail function menu”, press the “Operate” option b
111
4 1.1 Accidence of Panel and Display Information 1.1.1 Front Panel It is important for you to understand the DSO’s front panel before
112 Picture 2.13-21 Table 2-62 Waveform play back function menu 2: Option Setting Instruction Start Frame Set start frame. Curr_Frame
113 menu”. 6. Select the “
114 2.13.9 Recorder The waveform recorder is a kind of seamless and no-gap real time recording of waveform, means oscilloscope can sa
115 Table 2-67 recorder’s
116 Recorder Operation step: 1. Press the UTILITY button to enter “Utility menu”. 2. Press the “Next Page” option button to ente
117 2.14 Online Help Functi
118 Chapter3 Application Examples This section presents a series of application examples. These simple examples highlight the features of the oscill
119 3.1Taking Simpl
120 The corresponding icon and measure value will display on the third option area. ⑵⑵⑵⑵ Measure signal peak-to-peak amplitud
121 3.2 Taking Curs
122 3.2.2 Measuring Ring Amplitude To measure the amplitude, follow these steps: 1). Press the “CURSORS” button to see the Cursor Menu. 2). Press th
123 3.3 Catch the s
124 3.4 Analyze the signal detail You have a noisy signal displayed on the oscilloscope and you need to know more about it. You suspect that the sig
125 3.5 Triggering
126 3.6. Application of X-Y function Viewing Impedance Changes in a Network Connect the oscilloscope to monitor the input and output of the circ
127 DCA B P
128 3.7 Analyzing a Differential Communication Signal You are having intermittent problems with a serial data communication link, and you suspect p
129 Chapt
130 USB Flash Drive to the USB Host port. USB Flash Drive Pull Out! : This information will appear when you pull out the USB Flash Drive. Store
131 4.2 T
6 5. Power Shutoff Switch and Input Connector Picture 1.1-4 Back and Side panel connection locations on SDS1000C/CM/CE Seri
132 Check the trigger mode on the trigger interface whether or not in the “normal” or “single”, and check the trigger level is whether o
133 Chapt
134 5.2 Contact with us MTR Add : 3//F, Bldg No.4, Antongda Industrial Zone, 3rd Liuxian Road ,Bao’an District, Shenzhen, 518101, P.R.
135 Appendix A: Specifications All specification applies to 10X
136Probe attenuator Factors Set 1X、10X、100X、1000X Vertical System Vertical Sensitivity SDS1000C/CE/CF/CFL 2mV/div - 5V/div(1-2-5 order) SDS1000CM
137 <3.5ns (SDS1102C/CM/CE,SDS1104CF/CFL) <5.8ns
138 Channel Mode Samping Rate Short memory Long Memory Single Channel 1Gsa/s 40kpts No Support Single Channel 500MSa/s or lower 20kpts 2Mpts Me
139 X-Y Mode X-pole Input / Y-Pole Input Channel 1 (CH1)
140 Display System Display Mode Color TFT 5.7in.(145mm)diagonal Liquid Crystal Display Resolution 320 horizontal by 234 vertic
141 Gen
142 Appendix B: Default setup Menu or system options、、、、 knobs or buttons Default setup Coupling DC BW limit Off Volts/div Coarse Probe 1X
143 Persi
144 Type Alternative Source CH1 Mode Edge TRIGGER (Alternative) Coupling DC
145 Appen
146 Index - Width 75 + Width 75 Accessories 122 Acquiring Signals system 57 Aliasing 60 Application of X-Y function 107 AUTO 15 auto setup 1
8 indicates the filter is On. 11. Readout shows the main time base setting. 12. Readout shows the horizontal positon of main timebase waveform. 13.
10 1.3 Probe 1.3.1 Probe Safety A guard around the probe body provides a finger barrier for protection from electric shock. Picture 1.3-1
11 Be sure that the attenuation switch on the probe matches the Pr
II Declaration Copyright © by Siglent Technologies Co,.Ltd. All rights reserved. Contents in this Manual are not allowed to copy, extract and tra
12 Over Compensated under Compensated
13 Chapter 2 Functions Instruction and Operation To use your o
14 2.1 Menu and control button Showing as the following picture: Picture 2-1 All models: Channel buttons(CH1,CH2,CH3,CH4): P
15 SAVE/RECALL: Press to display the Save/Recall menu. You can
16 2.2 Connector Picture 2-2 Channel Connector (CH1,CH2,CH3,CH4): Input connectors for waveforms display. EXT TRIG: Input connector for an
17 2.3 Auto set
18 ● No signals found and no channels displayed, oscilloscope displays and uses channel 1. Table 2-2 Auto set the functio
19 2.4 Default setup The o
20 2.5 Universal knob Picture 2-5 Universal knob You can use the Universal knob with many funct
21 2.6 Vertical System T
III SDS1000 Series Brief Introduction Model: SDS1000C Series: SDS
22 Table 2-4 CH1,CH2 ,CH3,CH4 function menu 2: Table 2-5 Digital Filter function menu: Option Setting Instruction Invert on off Turn on i
23 “GND” Coupling: Use GN
24 Setting up CH1,CH2,CH3,CH4 Channels Each channel has its own separate Menu. The items are set up separately according to each chan
25 frequency component whic
26 Picture 2.6-4 4.Setting Probe Attenuation In order to assort the attenuation coefficient, you need to response in the channel
27 5. Inverting waveforms T
28 ● Press “CH1”→ “Next Page page1/2”→ “Filter” → “On”. Turn on the Digital Filter function. Picture 2
29 changes the Volts/Div sc
30 Picture 2.6-9 1. FFT Spectrum Analyzer The FFT process mathematically converts a time-domain signal into its frequen
31 Table 2-9 FFT function m
IV ● Cursor measure types: Manual mode, Track mode and Auto mode. ● Channel waveform and its FFT waveform display on split screen. ● Waveform Intensi
32 2. Displaying the FFT Spectrum Press the MATH button to display the Math Menu. Use the options to select the Source channel, Window algorithm, a
33 need to use. Table 2-
34 5. Measuring an FFT Spectrum Using Cursors You can take two measurements on FFT spectrums: magnitude (in dB) and frequency (in Hz). Magnitude is
35 Picture 2.6-11 2. Mea
36 2.6.4 Using Ref The reference control saves waveforms to a nonvolatile waveform memory. The reference function becomes availabl
37 Press the Ref button to
38 2.7 Horizontal System As follow Picture, there are one button and two knobs in the HORIZONTAL area. Pictu
39 MemDepth Detailed Speci
40 display updates from left to right. There is no trigger or horizontal position control of waveforms during scan mode. 2.7.2 Window Zone Use th
41 4. Press the “Window” b
General sa
42 2.8 Trigger System The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly,
43 sweep speed selected. 2.
44 Normal Use this mode when you want to see only valid triggered waveforms; when you use this mode, the oscilloscope does not display a waveform
45 2. Set up Source
46 When (Positive pulse width less than pulse width setting) (Positive pulse width larger than pulse width setting) (Positive pulse width equal
48 Picture 2.8-5 Table 2-18 Functional Manu of Video Trigger 2:
49
50 Table 2-19 Slope trigger function menu 1 Picture 2.8-7 Table 2-20 Slope trigger function menu 2 Option
VI ● Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present. ● Do Not Operate With Suspected Failures. If yo
52 6. Press the “When” option button to select “ ”, “ ”, “ ”, “ ”, “ ”或 “ ”. 7. Press the “Time” button, turn the “Universal” knob to adjust sl
53 Table 2-21 Set trigger m
54 Table 2-23 Set trigger mode to pulse trigger function menu 2: Table 2-24 Set trigger mode to video trigger function menu1:
55 Table 2-26 Set trigger m
56 4. Press the “Type” option button to select “Alternative”. 5. Press the “Channels” option button to select “CH1-CH2” 6. Press the “Source” opt
57 2.8.5 Slope & Level
58 Holdoff timeTrigger positionTrigger level Picture 2.8-11 If you want to change holdoff time, please follow next operations: 1. Press the “TRIG
59 2.9 Acquiring Signals
60 Sampling: In this acquisition mode, the oscilloscope samples the signal in evenly spaced intervals to construct the waveform. This mode accurat
61 Average: The os
General sa
62 the trigger point. This is also called the pre-trigger. 2). Continue to acquire data while waiting for the trigger condition to occur. 3). Detect
63 You can also turn Sins/x
64 2.10 Display System The display function could be expressed by the “DISPLAY” Button. Display System Picture 2.10-1 Table 2-2
65
66 Picture 2.10-3 Table 2-31 Display system function menu3: Operate Introduction: 1. Set up waveform display type 1) Pr
67 Pic
68 8. Set Skin Press the “skin” option button or turn the “Universal” knob to select “Classical”, “Modern”, “Traditional” or “Succinc
69 ● Turn the “S/div” knob
70 maximum values of a waveform and knew you had a scale factor of 100mV/div, then you could easily calculate your peak-to-peak voltage as follows:
71 parallel cursors to m
VIII Preface This manual contains operating information for the SDS1000 Series Digital Storage Oscilloscopes. The manual consists of the following
72 The reciprocal of time increment between Cursor A and Cursor B: 1/△T Picture 2.11-2 Track mode Table 2-33 Track
73 3. Press the “Cursor A”
74 Picture 2.11-3 Auto mode Table 2-34 Auto mode function menu: Option Setting Instruction Cursor Mode Auto Set to au
76 ROVShoot, FOVShoot, RPREShoot, FPREShoot parameter. 、 、 、 、 、 、 、 、 、 、 、 、 、 Display the corresponding icon and measure value of your s
77 Off measure voltage par
78 + Wid + Width Measures the time between the first rising edge and the next falling edge at the waveform 50% level. - Wid -Width Measures th
79 Picture 2.11-7 6 ) . P
80 2.12 Storage System Showing as following picture, The SAVE/RECALL is the Storage System Functional Button. Storage sys
81 While File shows option
82 Creating Folders and Files Creat new folders and files by pressing the New Folder or New File option button. The following screen is shown. Pict
83 SAVE/RECALL SETUP Sav
84 3). Press the “Save to” option button to select “Device”. 4). Press the “Setup” option button to select “No.1”. 5). Press the “DISPLAY” button to
85 Picture 2.12-7 Save se
86 Recall setup data from USB flash drive, follow next steps: 1) Press the “SAVE/RECALL” button. 2) Press the “Type” button to select “Setups”. 3) I
87 SAVE/RECALL WAVEFORM
88 want to need. 7). Press the “Save” option button. Recall wavefroms, follow next steps: 1). Press the “SAVE/RECALL” button to enter “SAVE/RECALL”
89 the oscilloscope
90 Picture 2.12-11 Save waveform image to USB flash drive, follow next steps: 1). Select the screen image that you want . 2) Press the “SAVE/RECALL
91 Save/Recall CSV