Service Manual
Chapter 1 Specifications
All specifications apply to 10X probe and SDS1000D Digital Storage Oscilloscopes. To verify
that the oscilloscope meets specifications, the oscilloscope must first meet the following
conditions:
●
The oscilloscope must have been operating continuously for thirty minutes within the
specified operating temperature.
●
You must perform the Do Self Cal operation, accessible through the Utility menu, if the
operating temperature changes by more than 5° C.
●
The oscilloscope must be within the factory calibration interval
①
①①
①
.
All specifications are guaranteed unless noted “typical”.
Specifications
Inputs
Input Coupling AC, DC, GND
Input Impedance 1MΩ±2% || 16Pf±3Pf
Maximum input voltage
400V (DC+AC PK-PK,1MΩ input impedance),CAT
I,CAT II
Probe attenuator
1X、10X
Probe attenuator Factors
Set
1X、5X、10X、50X、100X、500X、1000X
Vertical System
Vertical Sensitivity
2mV/div - 10V/div(1-2-5 order)
Channel voltage
offset range
2mV-200mV: ±800mV
206mV-10V: ±40V
Vertical Resolution 8 bit
Channels 2
Analog
Bandwidth
200MHz(SDS1202D)
150MHz(SDS1152D)